Read three numbers — rate, amplitude, beat error — and then read the fourth field, the lift angle. A reading taken at the wrong lift angle is a wrong amplitude wearing a decimal point.

How to read a timing machine

issue ISSUE 01checked 2026-09-16responsible Escapement Papers (a named bench, not a person)

RATEs/dthe periodAMPLITUDEdegreestrain healthBEAT ERRORmsthe two beatsLIFT ANGLEstatedmakes the rest usableTHE FOUR-CELL MEASURED CARD
The four-cell card. Three are readings; the fourth is the disclosure that makes them usable.

What you need A timing machine or a phone timing app; the movement's lift angle; the watch at a known state of wind; a note of the position being measured.

tape card: the three numbers and the fourth field

ReadingHealthy rangeWhat it tells you
Rate-4 to +6 s/dThe period of the oscillator
Amplitude250-310 degreesThe health of the train and lubrication
Beat error0.0-0.6 msHow evenly the two beats are spaced
Lift anglestated by youNothing on its own; it changes the other two

Steps

  1. Set the lift angle for the movement before you read anything. If you do not know it, say so and treat the amplitude as unusable.
  2. Wind the watch fully and let it settle for a few minutes. A reading taken on a half-wound watch measures the mainspring, not the movement.
  3. Read in the position you care about, and write the position down. Dial up and crown down are not the same watch.
  4. Read rate first, then amplitude, then beat error. If the amplitude is under about 200 degrees, the rate reading is unreliable and should be repeated after the watch has run.
  5. Repeat on a second position and compare. A rate that moves with position is a poise or a lever problem, not a regulation problem.

What each number is actually comparing

ReadingMeasured againstSensitive to
RateThe machine's own referenceOscillator period, regulator setting
AmplitudeThe beat waveform, via the lift angleTrain health, lubrication, mainspring
Beat errorThe gap between two successive beatsFork and roller clearance, banking
Lift angleNothing — you supply itEverything about the amplitude figure

The field nobody sets

Amplitude is not measured directly. The machine times how long the balance takes to cross the lift angle, and divides. If the lift angle you entered is wrong, the amplitude it prints is wrong, and the rate reading can shift slightly as well. Two machines with different default lift angles will disagree about the same watch while both being correct.

The bench holds that a published amplitude without its lift angle is not a measurement, it is a rumour, and that this should be the first question asked about any timing figure quoted online.

References

  1. Ref 1Measuring Instruments, Horopedia – Watchmaking Encyclopediasupports the instrument class a timing machine belongs to
  2. Ref 2Quality Testing, Horopedia – Watchmaking Encyclopediasupports the quality-testing framework these readings sit inside
  3. Ref 3NIST Time and Frequency Publication Databasesupports the published reference material a timing reading is compared against

One link, one document, titled as it titles itself. No home pages, no search results.

Set this page down Set this page down when you have written down the lift angle alongside the three numbers.